LOX-O2-F UV Flux 25% Oxygen Flow Through Sensor
The LOX-O2-F UV Flux 25% Oxygen Flow Through Sensor (formerly known as OX-0055) is designed to be used anywhere 0-25% oxygen levels need to be monitored in a closed system. This makes it a valuable tool for industry, researchers or scientific study.
The sensor benefits from lower power requirements than electrochemical and galvanic sensors, while providing a longer lifetime due to the non-depleting sensing principle.
The LOX-O2 is ppO2 and temperature compensated, enabling linear operation over a wide range without the need for additional system components. It does not contain lead or any other hazardous materials and has negligible cross sensitivity to other gases.
Features
- Flow-through housing for easy sample gas management
- Fast response T90 <10s
- Works in vacuum and high oxygen concentrations
- RoHs & REACH compliant
Models
- LOX-O2-F: UVFlux Oxygen Sensor – For research, science, experiments or integration into high-volume OEM products.
- CM-42990: UVFlux Oxygen Sensor w/ MX Board and Flow Through Tube Cap
- CM-42911: UVFlux Oxygen Sensor Development Kit
- CM-200: Sensor Development Kit - Easy to use, simply plug the sensor into the PCB and plug it into your PC via USB. Use our free GasLab® software to measure and graph carbon dioxide or use our library of source code on Github to build your own application or for OEM batch testing/configuration. Order the sensor on this page, then order the Devkit here.
We recommend that all first time users purchase both this sensor and the CM-200 Development Kit for fastest and easiest results.
Specifications
- Measuring Gas: Oxygen
- Measurement Range: 0 - 25% O2
- Sensing Method: Fluorescence quenching by oxygen
- Measurement Interval: 1 Sample/second
- Oxygen Pressure Range: 0 — 300mbar ppO2
- Response Time: T90 < 10s
- Warm-Up Time: < 60 Seconds at 72°F (22°C)
- Accuracy: Better than 2% at full scale
- Humidity: 0-99% RH (non-condensing)
- Resolution: 0.10% / 0.1mbar
- Temperature Accuracy: +/- 2 degrees C
- Pressure: ±5mbar
- Operating Temperature: -30°C to +60°C
- Barometric Pressure Range: 500 to 1200mbar
- Lifetime: > 5 years
Electrical / Mechanical
- Supply voltage: 4.5 - 5.5 VDC
- Supply current: <7.5mA (streaming one sample per second)
- Output type: 3.3V TTL level USART
- Interface: UART or analog
Warranty
- 90 days - See CO2Meter Terms and Conditions.
- Gas Sensor Development Kit (CM-200)- designed to provide customers with an easy drop-in solution. Includes our free Gaslab data logging software. Sensor sold separately.
- Micro-Pump Kit includes all the tools you need to utilize the tube-cap adapters as part of a sampling sensor system.
- Nafion Tubing is a highly selective, semi-permeable membrane to water vapor. If gases inside Nafion tubing are wetter than gases surrounding the tubing, drying will occur. If the surrounding gases are wetter, humidification will occur.
- Micro Gas Sampling Pumps provide a constant flow of air across the sensor with a tube cap - perfect for bench-top testing of sensors for closed-loop sampling applications.
-