How Does an Oxygen Sensor Work?

How Does an Oxygen Sensor Work?

How does an O2 sensor detect oxygen?

What is an oxygen sensor? How do they work? While there are many types of oxygen sensors, their working principle can be categorized in one of 3 ways:

  1. A chemical reaction that gives off electrons in the presence of oxygen.
  2. A change in light intensity given off by a fluorescing material when exposed to oxygen.
  3. A change in the wavelength of sound, light or in a magnetic field as oxygen passes through it.

Each of these ways to measure oxygen has strengths and weaknesses. While oxygen sensors are used in many applications and industries including automotive, health and medicine, industrial, food and beverage packaging, pharmaceutical and more,, each uses a different type of oxygen sensor best suited for the application.

Note that most oxygen sensors are designed to measure between 0 and 25% oxygen by volume or in breathable air. However, specialized oxygen sensors that can measure up to 100% oxygen are also available. 

It is important to note that the oxygen sensor does not actually measure oxygen concentration, but rather the difference between the amount of oxygen in the exhaust gas and the amount of oxygen in the air. 

What are the different types of oxygen sensors?

  1. Electrochemical oxygen sensor
  2. Zirconia oxygen sensor
  3. Optical oxygen sensor
  4. Clark oxygen sensor
  5. Infrared oxygen sensor
  6. Electro galvanic sensor
  7. Ultrasonic oxygen sensor
  8. Laser oxygen sensor
  9. Paramagnetic oxygen sensor

Shop O2 Monitoring Solutions

Below are the specific types of oxygen sensing technology used today. Note that each one is best suited for one or more specific applications. 

1. Electrochemical Oxygen Sensor

electrochemical oxygen sensor

Electrochemical oxygen sensors are primarily used to measure oxygen levels in ambient air. They measure a chemical reaction within the sensor that creates an electrical output proportional to the oxygen level. Because electrochemical sensors produce a current, they can be self-powered, making them useful for measuring oxygen gas battery-operated underwater diving and hand-held personal safety devices. Examples include breathalyzers, respiratory sensors, and blood glucose sensors.

In terms of sensor advantages, electrochemical sensors are sought after due to their low power requirements, lower detection limits, and are often less directly affected by interfering gases. They also tend to be the least expensive kind of sensor.

A challenge for electrochemical oxygen sensors is that they depend on a chemical processes that is temperature dependent. The output of most electrochemical sensors will rely heavily on temperature compensation to provide reliable readings over a broad scope of ambient conditions.

Another challenge for electrochemical oxygen sensors is that over time the chemical reaction slows down and stops, typically between 1 and 3 years depending on the sensor design. Storing them in an oxygen-free environment will not add to the life span of the sensor. As the sensor ages it requires frequent recalibration and is not as accurate as other sensors.

However, because of their rugged design, low cost, and self-powering, electrochemical oxygen sensors are used in many devices, especially hand-held gas analyzers.

AlphaSense is one of the most popular manufacturers of electrochemical oxygen sensors. Their sensors are used in dozens of multi gas detectors and portable safety meters used worldwide.

2. Zirconia Oxygen Sensor

Zirconia oxygen sensors use heat and chemistry to detect oxygen. Zirconia dioxide is coated with a thin layer of porous platinum to form a solid‐state electrochemical fuel cell. Carbon monoxide, if present in the test gas, is oxidized by the oxygen to form CO2 which triggers a proportional flow of current. The zirconia sensor does not directly sense O2, but rather the difference between the concentration of oxygen in the sample gas and in fresh air.

zirconia oxygen sensor

While zirconia oxygen sensors are most commonly used to control air-fuel ratios in cars and trucks, they are also important in industrial applications. For example, SST’s Zirconia Oxygen Measurement Sensor System uses this technology to measure the oxygen content in flue gases, combustion control systems, coal, oil, gas, biomass, and oxygen generation systems

Another feature of this type of oxygen sensor is that the small, zirconium-based element does not require calibration. They also maintain their accuracy, even when exposed to humidity or other gases.

Because of a zirconia oxygen sensor's ability to work at high temperatures and pressures, the possible application make it useful in the automotive industry. Virtually every car or truck manufactured uses two zirconia oxygen sensors, also known as lambda sensors, to adjust the fuel-air ratio to maximize combustion efficiency.

The disadvantage of zirconia sensors is that oxygen measurement requires high temperatures. During use, a heater in the sensor raises the sample gas to above 300°F. The heater needs lots of power, so zirconia oxygen sensors are not used in battery operated or handheld devices. In addition, zirconia sensors are not useful where very high accuracy is required.

A variation on the zirconia oxygen sensor is the planar oxygen sensor. Like a traditional zirconia oxygen sensor, it is moisture resistant, rugged, and requires  a built-in heater to operate. However, instead of zirconia it uses alumina, which is able to achieve the required temperature more quickly. As a result, a planar oxygen sensor can begin reading oxygen levels in less than 10 seconds instead of the normal 30 second warm up time of a traditional zirconia sensor. This advance makes it a better alternative to automotive lambda sensors for reducing NOX gases present during cold starts.

3. Optical Oxygen Sensor

Optical oxygen sensors are based on the principle of fluorescence quenching by oxygen. They rely on the use of a light source, a light detector, and a luminescent material that reacts to light. In many fields, luminescence‐based oxygen sensors are replacing the Clark electrode.

The principle behind fluorescence quenching by molecular oxygen has long been understood. Some molecules or compounds, when exposed to light, will fluoresce (i.e. emit light energy). However, if oxygen molecules are present, the light energy is transferred to the oxygen molecule resulting in less fluorescence. By using a known light source the amount of light energy detected is inversely proportional to the number of oxygen molecules in the sample. Therefore, the less fluoresce detected, the more oxygen molecules must be present in the sample gas.

In some sensors, the fluorescence is detected twice at a known time interval. Instead of measuring the total fluorescence, the drop in luminescence (i.e. fluorescence quenching) over time is measured. This decay-based time method allows for a simpler sensor design.

An example of a sensor that measures ambient oxygen levels using fluorescence quenching by oxygen is the LuminOX LOX-02 sensor. While it has the same footprint as traditional electrochemical sensors, it does not absorb oxygen and it has the advantage of a much longer lifespan. This makes it useful for devices like room oxygen depletion safety alarms which monitor indoor air for a sudden drop in oxygen levels from stored compressed gases.

Common applications that involve optical sensors include medical facilities, lasers, imaging systems, and fibers. In regards to sensor advantages, many find optical sensors to hold greater sensitivity, wider dynamic range, distributed configuration and multiplex capabilities.

oxygen sensor

Another example is the TecPen Modified Atmosphere Packaging Handheld Oxygen Analyzer. The TecPen uses a thin coating of luminescent dye on the sensor and a micropump to pull the air sample past the fluorescing dye. The dye is excited at 507 µm and the resulting fluorescence event recorded at 650 µm. The duration of this fluorescence event – known as the lifetime – depends on the quantity of adsorbed oxygen in the sensor layer and can thus be used to determine the oxygen concentration.

Because it uses the faster optochemical sensing technology it is able to take a measurement in seconds. In addition, optical oxygen sensors can be very accurate with the ability to measuring oxygen in the parts per billion level. This makes optical oxygen sensors useful in processes like modified atmosphere packaging or weld purge monitoring that need to measure the absence of oxygen down to 3-4 parts per billion oxygen molecules.

4. Clark Electrode Oxygen Sensor

Blood Glucose Testing

The Clarke electrode is a type of electrochemical oxygen sensor. It measures oxygen levels in liquid using a cathode and an anode submerged in an electrolyte.

The Clark electrode was invented to measure oxygen levels in the blood during cardiac surgery. Today it is commonly used in portable blood glucose monitoring devices that require a drop of blood.

The sensor uses a thin layer of glucose oxidase (GOx) on an oxygen electrode. By measuring the amount of oxygen consumed by GOx during the enzymatic reaction with the glucose, the blood glucose level can be calculated and displayed.

Additional Clarke-type sensors are available which include measuring of ozone (O3), Hydrogen Peroxide (H202), Hydrogen (H), and Hydrogen Sulphide (H2S).

While only accurate to tenths of a percent of oxygen, their low cost has made Clarke electrode oxygen sensors available as consumer products.

5. Infrared Oxygen Sensor

Wrist-oximeter

Infrared pulse oximeters, commonly called fingertip oximeters or finger pulse oximeters, are oxygen sensors that measure the amount of oxygen in the blood by light. They are most often used in low-cost fingertip or earlobe devices to measure oxygen saturation in the body for home medical use.

To work, infrared and red light are both pulsed through a thin layer of skin and measured by a photo diode. Because the wavelengths of the 2 light sources are different, the ratio of absorption of light through the skin is proportional to the amount of oxygenated hemoglobin in the arteries.

The advantages to purchasing infrared oxygen sensors are due to the fact that they are noninvasive, cost-effective, compact and easily can quickly detect low oxygen levels in the blood. Their downside is that some of the less expensive models are not approved as medical devices due to low accuracy and repeatability.

6. Electro Galvanic Sensor

electro galvanic Sensor

An electro-galvanic oxygen sensor is a fuel cell based on the oxidation of lead that produces an electrical output proportional to the oxygen level inside the sensor. It is similar to an electrochemical sensor in that it consumes itself over several months as it is exposed to oxygen.

Because electro galvanic sensors are relatively low-cost and dependable devices that can measure 0-100% oxygen levels, they are used as medical oxygen sensors in many hospital ventilators as well as SCUBA diving equipment. The downside of electro galvanic oxygen sensors like medical oxygen cells is that they typically have a lifespan measured in months. These sensors tend to be accurate within tenths of a percent of oxygen.

7. Ultrasonic Oxygen Sensor

HC SR04 Ultrasonic sensor 1480322 3 4 HDR Enhancer

Ultrasonic oxygen sensors use sound speed to measure the amount of oxygen in a gas or liquid sample. In liquid, upstream and downstream sensors measure the velocity difference between high-frequency sound waves. The change in velocity is proportional to the amount of oxygen in the sample. In gases, the sound speed varies as the molecular composition of the gas varies. This makes ultrasonic oxygen sensors useful for anesthesia ventilators or oxygen generators where the output is a known concentration of oxygen gas. Typical applications that require ultrasonic oxygen sensing methods are hospitals, gas analysis, or applications involving oxygen concentrators or portable oxygen generators. 

8. Laser Oxygen Sensor

Tunable laser diode oxygen sensor

Tunable Diode Laser (TDL) oxygen sensors rely on spectral analysis. A laser beam at the wavelength of oxygen is directed through a gas sample to a photodetector. The amount of light absorbed by the oxygen molecules is proportional to the number of molecules in the sample.

The mechanism of the laser oxygen sensor was created to design analyzers for real-time measurement of gases such as H20, H2S, CO2, NH3, and C2H2 in gas streams. Many sensors have been used in various applications such as combustion systems, power plants, coal, and waste incinerators.

The benefits of laser oxygen sensor is their fast response time, accuracy within tenths of a percent oxygen, that they are inherently calibration-free and their long life. Their disadvantages are primarily their susceptibility to cross sensitivity from other gases.


Sources:

https://onlinelibrary.wiley.com/doi/full/10.1002/bies.201500002

https://o2sensors.com.au/static/what-is-oxygen-sensor

https://www.newswire.com/different-types-of-o2-sensors/23890

https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4744989/

https://www.systechillinois.com/en/support/technologies/paramagnetic-cells

https://en.wikipedia.org/wiki/Electro-galvanic_oxygen_sensor

https://pubs.acs.org/doi/10.1021/acssensors.2c00938

Image by pixabay


Article précédent Article suivant