Tecpen MAP Oxygen and Carbon Dioxide Analyzer
The Tecpen MAP Oxygen and Carbon Dioxide Analyzer ensures accurate oxygen and carbon dioxide gas level measurements in pierceable, shrink-wrapped packaging.
By using the TS-202 individuals are able to gain precise O2 and CO2 measurements in just seconds while safely monitoring perishables, ensuring food protection, and mitigating the risk of potential food-borne illnesses.
Features
- Accurate O2 optochemical sensor
- Long-life NDIR CO2 sensor
- Measurements in 3 seconds for oxygen, 10 seconds for CO2
- Data can be logged and downloaded via USB
- Industry-standard .CSV data files
- Rechargeable battery up to 300 measurements per charge
- Certificate of calibration from the factory
- Not cross-sensitivity to carbon monoxide, hydrocarbons
- No cross-sensitivity to H2S, ionic strength, flow
Applications
- Modified atmospheric packaging (MAP) for Quality Control
- Pharmaceuticals and biotechnology
- Hygiene detection (pathogenic), microorganism growth
- Biochemical and molecular biology research
- Cell culture vessels, Infusion containers
- Micro-bioreactors
Models
Specifications
Oxygen Sensor
- Measurement Range: 0-2,000ppm
- Accuracy @ 0-0.5%: ± 2%Mev**
- Accuracy @ 0.5-2.5%: ±3% Mv*
- Accuracy @ 2.5-5%: ±5% Mv*
- Resolution: 1ppm
- Response time at 25°C/ 77°F: <150ms
- Pump flow: 400mL/min.
- Temp. range Min./Max: -10°C/ +60°C (14°F/140°F)
- Temperature compensation: 10-30°C (50-86 °F)
- Zero-point adjustment interval: adjusted to air at power on
CO2 Sensor
- Measurement Range: 0-5%
- Accuracy @ 1%: ±70 ppm +/- 5% of measuring value
- Accuracy @ 100%: ±300 ppm +/-5% of measuring value
- Resolution 1ppm
- Pump flow: 400mL/min.
- Response time at 25°C/ 77°F: 10s
- Temp. range Min./Max: -25°C/55°C (-13°F/ 131°F)
- Temperature compensation: 10-30°C (50-86 °F)
- Recommended zero-point adjustment interval by customer: 6 months
Device
- Pump flow max.: 67 mL
- Medium: Gas
- Power supply: 5V USB and LiPo battery
- Rechargeable battery lifetime: > 3 hours per charge
- Data Interface: USB
- Display: OLED
- Cleaning housing: no organic solvents. 40% EtOH
- Parts touching sample: St.1.4404/ PTFE/ Glas
- Case: Anodized aluminum
- Protection Class: IP65
- Full-service interval by Manufacturer (includes necessary retrofitting): 3 years
- Life time of O2 sensor (protective atmosphere; 0- 40°C): 2 years
- Sensor cap replacement interval: 2 years
- Case with Device Weight: 2.45 lbs.
- Case Dimensions: 12 x 11 x 3.25 inches
- Device Weight: 0.50 lbs.
- Device Dimensions: 7 x 2 x 1 inch
- **Mev = measured end value
*Mv = measured values
Please note: TecSense sensors are CE, FCC, IC, KC, and MIC marked, meet the requirements of ASTM F2714-08, and the European and Canadian standards for food safety. The company is DIN ISO 9001/2015 certified.
Warranty
Accessories
- PRT-SEPTUMS - septum's (pack of 18) - contact us
- PRT-NEEDLES - 25/0.8 needles (pack of 10) - contact us
- CM-0118 - hydrophobic filter qty. 10 - contact us
- TS R1 HH 007 - flexible extension - contact us
-
Data Sheets
-
Manuals
-
Technical Documents
-
Videos
-
Additional Resources
-
Case Studies